Page view 1
SPECIFICATIONS
Analog Inputs
8 analog inputs:
Thermocouple J, K, T, E, N, R, S and B (IEC-584).
Pt100, Pt1000 (IEC-751), 0-20 mA, 4-20 mA, 0-
20 mV, 0-50 mV, 0-60 mV, - 20-20 mV, 0-5 V
and 0-10V.
Analog Accuracy
Thermocouple J, K, T, E and N: 0.2 % of range ±
1 °C.
Thermocouple R, S and B: 0.2 % of range ± 3 °C.
Pt100, Pt1000, 0-20 mA, 4-20 mA, 0-20 mV, 0-50
mV, 0-60 mV, - 20-20 mV, 0-5 V and 0-10 V: 0.15 %
of maximum range.
Digital Inputs /
Outputs
8 digital channels that can be individually configured
as inputs or outputs.
Relay Outputs
2 Relays SPST-NO/NC, 3 A / 250 Vac
RS485 Interface
Can be used as a Modbus RTU slave or master (can
read up to 64 slaves).
Ethernet Interface
10/100 Mbps.
Supported
Protocols and
Services
TCP/IP, DHCP, HTTP, SMTP, SNMP, Modbus RTU,
Modbus TCP, FTP server and client.
Internal Memory
2 MB.
SD Card
Up to 16 GB.
Operating
Environment
0 to 50 °C. Relative Humidity (Maximum): 80 % up to
30 ºC. For temperatures above 30 ºC, decrease 3 %
per ºC. Installation category II. Pollution degree 2.
Altitude < 2000 m.
Enclosure
ABS + PC
Dimensions
165 x 117 x 70 mm; Weight: 400 g. (no HMI)
Power Supply
100 to 240 Vac/dc (50 / 60 Hz)
24 Vdc/ac (50 / 60 Hz)
Protection
IP20
Configuration
Software Operation
Environment
Software for Windows® XP, Vista and 7.
Menus in Portuguese, Spanish or English.
Table 01