Novus DC80L User Manual Page 11

  • Download
  • Add to my manuals
  • Print
  • Page
    / 13
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 10
11
DC mV and V generation
1. Press the power switch to
turn on the Calibrator.
2. Press the INPUT / OUTPUT
mode key
OUTPUT mode .
3. Press the V / mV mode key
scale. The mode indication (V or
mV) is shown in
.
4. Use the increment / decrement
keys
to adjust
the output to the desired value.
5. Plug the red test lead in the “V
jack
and the black one to
the COM jack
.
6. C
(red) and the negative test lead
(black) to the target equipment.
7. For changing the output value,
use the increment / decrement
keys
and the V
/ mV mode key
until the
desired value is displayed.
DC mA measurement
a) Using external power supply
1. Press the power switch to turn
on the Calibrator.
2. Press the INPUT / OUTPUT
selection key to select input
mode. The INPUT sign is shown
on the display
.
3. Press the mA / mA % mode key
, to select mA or mA % .
4. Plug the red test lead in the mA
jack
and the black one to
the COM jack
.
5. Connect the test lead according
to the signal polarity (red to the
positive, black to the negative).
6. The measurement is shown on
the display
.